SABER UCV >
1) Investigación >
Artículos Publicados >
Por favor, use este identificador para citar o enlazar este ítem:
http://hdl.handle.net/10872/4650
|
Título : | An in-tube SPME device for the selective determination of chlorophyll a in aquatic systems |
Autor : | Vitta, Yosmery |
Palabras clave : | Chlorophyll a In-tube solid phase microextraction Capillary LC Water quality |
Fecha de publicación : | 28-Oct-2013 |
Citación : | Talanta 82;82 |
Resumen : | We report a new device for the estimation of the content of chlorophyll a pigment in water samples as an
indicator of water quality. The extraction of the pigment from water was also optimized. 10mL of water
was filtered through a nylon filter (45 mpore size and 13mmof diameter), after the chlorophylls were
dissolved by immersing the filter in 1mL of a low non-hazardous solvent as ethanol. An in-valve in-tube
SPME device coupled to capillary liquid chromatography with diode array detection was designed. A
capillary column of 70cm in length (0.32mm i.d. coated with 5% diphenyl–95% polydimethylsiloxane,
3 m coating thickness) was used as the loop of the injection valve for preconcentration and a Zorbax
SB C18 (SiO2-based) 150mm×0.5mm i.d., 5 m column (Agilent) was used as analytical column. The
achieved detection limit was 0.05 gL−1 and the working range of concentrations was 0.1–1 gL−1. %
RSD values between 2 and 11 were obtained. Chlorophyll a in several water matrices was determined
with good results in presence of other pigments such as chlorophyll b, pheophytin a and pheophytin b. |
URI : | http://hdl.handle.net/10872/4650 |
ISSN : | 0039-9140 |
Aparece en las colecciones: | Artículos Publicados
|
Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.
|